Enlarge Image
BariKare Therapy System
A front-exit therapy system for bariatric patients
The KCI BariKare™ Therapy System, for patients weighing up to 850 lbs., helps prevent skin breakdown and manage the challenges associated with bariatric patient care. The accompanying AtmosAir™ is a non-powered, pressure redistribution MRS with Self-Adjusting Technology and a sloped heel section to off-load pressure. Add the First Step Select™ Heavy Duty Overlay, for patients weighing up to 650 lbs., to help manage skin microclimate.
Features/Benefits
The BariKare Therapy System offers features and benefits designed to help treat and prevent pressure ulcers while managing the risks associated with bariatric patient care:
- Self-Adjusting Technology (SAT™) automatically and continuously adjusts to patient movement using non-powered air cylinders for maximized weight displacement and minimized tissue interface pressures
- Optimized Pressure Redistribution Therapy for patients over 500 lbs using SAT weight dials
- Heel protection with visco-elastic foam and a sloped heel section reduce tissue interface pressures by off-loading the heel
- Cardiac chair positioning facilitates up-in-chair orders, which enhance pulmonary positioning
- 850 lb. weight capacity
- Front exit facilitates patient ambulation
- Easy-to-use bed controls facilitate flexible patient positioning, allowing caregivers to raise and lower the bed and tilt the chair position as needed
- Built-in scale display provides instant readings at the touch of a button, located on both side rails
- Expandable side rails accommodate wider patients, while allowing easier patient transport through doors by rotating down and tucking against frame
- X-ray cassette holder eliminates patient transfers
- Fluid resistant and anti-bacterial mattress cover protects the mattress and allows for simple wipe down cleaning
Product Specifications
Frame and Mattress Dimensions
Overall: 87.5”L x 21.25” - 27.5”H
Mattress: 80”L x 36”W x 7”H
Side rails (Normal/extended/tucked): 43.25”/54”/40.25”
Side rail: 21.5”H x 48”L
Weight
Maximum recommended patient weight: 850 lbs.
Net weight: 665 lbs.
Articulation
Fowler: 65°
Leg Drop
Trendelenburg: +10°
Reverse Trendelenburg: -12.5°
Cardiac Chair
Fowler: 65°
Trendelenburg: +10°
Reverse Trendelenburg: -12.5°
Electrical
Volts: 110 / 115 V
Amps: 4.5
Maximum electrical leakage: <100 Microamps
Bed Frame: UL Listed
Flammability
All mattresses meet TB 106, TB 117, TB 603 standards
TB 129 and BFD IX-II Compliant
Scale
Scale repeatability: 99.3%
Scale accuracy: +/- 0.7% patient weight
Warranty
One year
Part Number
Complete assembly: 306000
Specifications are subject to change without notice.
Product Usage
Indications
- For patients whose conditions require med/surg features
- For patients whose body weight or size pose a significant risk for care management issues to the patient or staff during the performance of routine nursing care
- For the prevention and treatment of pressure ulcers
- For patients weighing up to 850 lbs. (including accessories)
Contraindications
- Unstable cervical, thoracic and/or lumbar fracture
- Cervical and skeletal traction
- Total weight in excess of 850 lbs. (including accessories)
Product Literature
Also available for download is a PDF version of the following:
BariKare Therapy System Brochure
First Step Select Heavy Duty Overlay Brochure
Please see product labeling and instructions for use for risks, precautions, warnings and detailed operating instructions prior to use.