Therapy Adjustment Options
Negative Pressure: 50-200 mmHg
Mode of Operation: Instillation, Continuous and Intermittent †
Intensity: 10-50 mmHg/second
Physical Data
Dimensions w/o stowed Instillation Pole: 15.6”(40cm) W x 11”(28cm) H x 7.1”(18cm) D
Dimensions with stowed Instillation Pole: 16.6”(42cm) W x 11”(28cm) H x 7.1”(18cm) D
Weight with Instillation Pole: 14.9 lbs. (6.8 kg)
Weight without Instillation Pole: 14.5 lbs. (6.6 kg)
Canister Volume: 500 ml
Battery Life: Approximately 4 hours
Electrical Data
Voltage:100-240 V~
Frequency: 50-60 Hz
Maximum Power Consumption: Approximately 70 Watts
Classification
Type B Applied Part.
Class II Equipment (double insulated) with internal electrical power source.
Classified by Underwriter Laboratory to UL 2601-1 and IEC 60601-1 (no leakage current test required).
Disposables are latex-free.
Specifications are subject to change at any time without notice.
In the interest of safety and efficacy, the V.A.C.® Instill® System is only for use with V.A.C.® Disposables, and V.A.C.® Disposables are only for use with V.A.C.® Therapy units.
V.A.C.® Instill® Pads are not packaged with T.R.A.C.® Dressings, only separately.
† Intermittent operation is only available when using standard V.A.C.® Therapy. Default therapy mode is continuous during Instill Therapy™.
Back to top
IMPORTANT: As with any prescription medical device, failure to consult a physician and carefully read and follow all therapy unit and dressing instructions and safety information prior to each use may lead to improper product performance and the potential for serious or fatal injury. Do not adjust therapy unit settings or perform therapy applications without direction from and/or supervision by your clinical caregiver.